Experimental apparatus
The goniophotometer used at Wellman Laboratory is shown in Figure
5.2. The interior of the tank was painted flat black and filled
with saline to minimize internal reflections from the box and
within the sample. A helium-neon laser delivered a 1mW
beam normal to the sample which was sandwiched between glass
microscope slides. The beam diameter was 1mm. The
3mm diameter detecting fiber located at the end of an
8.5cm long arm was attached
to a computer-controlled stepping motor. The stepping motor
made 1.8
steps throughout a full 360
circle.
A 1.17mm aperture was placed over the end of the detecting
fiber to increase the resolution of the goniophotometer. Light
collected by the fiber was measured with a photomultiplier tube
whose output was connected to the computer.
Figure 5.1:
Correction factor as a function of angle required
to convert measurements of reflection and transmission into phase
function measurements. These correction factors are for an optical
depth of unity and an albedo of one-half. The solid line is
the correction factor required for transmitted light and the
dashed line is that needed for reflected light.
|
|
Figure 5.2:
The goniophotometer apparatus. Helium-neon laser
light enters the saline filled tank through a glass port on the
side. The light strikes the sample and is scattered. The light
is detected by a fiber attached to an arm driven by a computer
controlled stepper motor. The fiber is connected to a photomultiplier
tube which is monitored by a computer.
|
|
The angular resolution
of the goniophotometer depends
on several different factors: the width of the detector
(Wdetector),
the acceptance angle of the detector
,
the width
of the beam on the sample
(Wbeam), and the distance from the
sample to the detector (D). The angle subtended by the beam
is approximately (Figure 5.3A)
 |
(5.5) |
Any light leaving the sample at angles larger than this will
not reach the detector. The angle subtended by the detector
is approximately (Figure 5.3B)
 |
(5.6) |
Again any light leaving the sample at angles larger than this
will not reach the detector. The angular resolution of the goniophotometer
is usually determined by the larger of
and
since
the acceptance angle of the detector is usually much larger than
the other angles. Thus, the larger of diameters of the beam
and the detector determine the resolution of the goniophotometer.
The angular resolution of the goniophotometer used for these
experiments was about 0.8 .
Figure 5.3:
Angular resolution of the goniophotometer is determined
by the relative size of the detector and the spot size when the
detector is sufficiently far from the sample. This figure defines
the angles subtended by the spot and the detector.
|
|
|