Evaluation of the method
The method for measuring phase functions was evaluated using
theoretical values for reflected and transmitted light from slabs
of varying thicknesses. The light exiting a uniformly illuminated
slab as a function of angle was calculated using the adding-doubling
method. All calculations assumed an albedo of 0.98 (the measured
albedo of the dermis at 632nm) and an isotropy factor of
gHG=0.9
(again, close to that of dermis at 632nm). Four sets of data
were calculated: matched and mismatched boundaries with
and .
Each of the four cases was calculated for a series
of optical depths ranging from
to .
The mismatched
boundary case was equal to that for tissue-glass-saline and the
indices of refraction were 1.37/1.54/1.33.
The reflected and transmitted light as a function of angle for
and ,
with matched and mismatched boundaries is shown
in Figure 5.7. Both boundary conditions manifest a strong forward
peak at 0 .
However, when the indices of refraction are
mismatched, a backward peak at 180
results from internal
reflection ( 1%) of the forward peak. The method for obtaining
the phase function from measured quantities was tested using
the four sets of calculated data. The known values of
and
gHG=0.9 are compared with the fitted values using the procedure
outlined in Section 5.4. The fitted parameters
and
gHG are
presented in Figures 5.8 and 5.9 as a function of the optical
thickness of the theoretical sample.
The values of
gHG for all four sets of data were identical. Figure
5.8 shows
gHG for only one set of data is shown ( =0, mismatched
conditions). This suggests that
gHG is insensitive to changes
in the isotropy factor
and boundary condition. For sample
thicknesse less than one optical depth the error in the
fitted phase function is less than five percent. Thicknesses
greater than one optical depth multiply scatter and cause large
errors in the fitted phase function.
Figure:
Light leaving a uniformly illuminated slab as calculated
by the adding-doubling method with a=0.98,
=0.9, .
Matched
(empty circles) and mismatched (1.37/1.54/1.3) conditions are
shown. The peak at
in the mismatched case arises
from internal reflection of the forward peak (
|
|
Figure:
Dependence of fitted anisotropy factor
on sample
thickness. The true value of
for all optical depths is 0.9.
The deviation of the fitted values from the true value is caused
by multiple scattering.
|
|
Figure 5.9:
Sensitivity of
to sample thickness and boundary
conditions. The empty points correspond to fitted values with
=0.1 and the filled correspond to =0. The squares
indicate mismatched conditions and the circles correspond to
matched conditions.
|
|
Figure 5.9 shows the dependence of the fitted isotropy parameter
as a function of optical thickness. The isotropy factor
is more sensitive to thickness than
gHG and errors do not become
small until optical thicknesses approach 0.1. Moreover,
is
very sensitive to boundary conditions and significant errors
exist for even very thin tissues with mismatched boundaries.
The discrepancy occurs because internal reflection increases
the path length of light in the tissue for which the single scattering
model cannot account The data presented in Figure 5.9 suggests
that
cannot be measured reliably because any real experiment
will have mismatched boundaries and subsequent analysis will
overestimate the isotropy factor. Therefore, Figure 5.9 should
be used to correct any measured value of
to obtain the true
value of
for light scattered isotropically.
|