omlc

Home

contents

up

next

previous

Light Transport in Tissue


Bloodless human dermis as a function of wavelength

A sample of human dermis was obtained from the abdomen at autopsy. The epidermis was manually separated following mild thermal treatment (two minute exposure in a 55$^\circ $C water bath). The dermal sample was soaked in saline prior to measurement to remove residual blood. One 2 x 2 centimeter sample was cut with a dermatome. The sample was sandwiched between glass microscope slides and the tissue thickness (360$\mu $m) was determined with micrometer.

A spectrophotometer (Beckman UV 5270) was used for measurements of reflection and transmission. The sample was placed at the entry port of the integrating sphere for measurements of total transmission. Collimated light directly struck the sample, and all light passing through the sample was collected by the integrating sphere. Diffuse reflection was measured by placing the sample in the exit port of the sphere. Specularly reflected light from the sample travelled back along the incident beam path and was not collected by the integrating sphere. For calibration purposes, zero and 98% reflection were obtained by measuring the reflectance with the sample removed and with a BaSO4 plate repectively. Collimated transmission measurements were made by removing the integrating sphere assembly and placing the sample in the path of the beam. Collimated light struck the sample, but only light propagating co-linear with the incident beam was detected. The measurements of reflection and transmission as a function of wavelength are presented in Figure 6.6.

The iteration algorithm was used to convert these measurements to optical properties. These properties are shown in Figures 6.7. Both absorption and scattering coefficients decrease with increasing wavelength. This indicates that longer wavelengths of light penetrate deeper into a tissue. The anisotropy increases with wavelength indicating that light scattering increases in the forward direction.

Figure 6.6: Total transmission, collimated transmission, and diffuse reflection from a 360$\mu $m sample of bloodless human dermis. Transmission increase with wavelength and reflection decreases with wavelength.
\includegraphics [scale=0.907]{fig66.eps}

Figure 6.7: Optical properties of bloodless human dermis as a function of wavelength.
\includegraphics [scale=0.907]{fig67.eps}

S. A. Prahl."Light Transport in Tissue," PhD thesis, University of Texas at Austin, 1988.